One
of the highest expenses incurred by Echocardiography Labs
is phased array transducer replacement (all major manufacturers)
due to CW Doppler noise.
Cause 
Technologists, lacking a drop leaf table or cut out mattress,
force transducers into the space between the patients apical
access window and the bed or table. This results in frequent
and radical flexing of the transducers bend relief. Within
a relatively short period of time the fine wires found in
newer light gauge cables break and short resulting in the
subject noise.
|
|
For
most users, the cost of remedying this problem is an expensive factory
transducer exchange.
Fix
CVS offers an affordable transducer exchange solution. We
re-terminate existing cables and replace the bend relief with
a longer and more rigid rubber compound, extending the future
life of our repaired transducers. |
|
Also, for labs not currently equipped with dropleaf tables or cut-out
mattresses, CVS offers inexpensive mattresses and a competitive
line of quality dropleaf tables.
Prevention

When a dropleaf table or cut-out mattress is not available,
avoid severe flexing of the bend relief with patient positioning
techniques, or use of CVS's "Wedgy" in hospital
portable applications, or Philips-Agilent (HP) Sonos 4500/5500
snap-on bend relief support.
Acoustic block replacement
Cable re-termination
Lens replacement
Bend relief replacement |
Housing and cap replacement
Cable patching
Connector pin & key replacement |
All
CVS repairs are carefully tested to insure they meet OEM specifications
and are fully warranted.

|
|